Recognition and diagnosis1Patient–ventilator asynchrony is a mismatch between the patient’s neural breath and the ventilator-delivered breath in timing and/or level of assist during assisted ventilation.
2In an NIV-focused definition, patient–ventilator asynchrony occurs when there is >20% mismatch between pneumatic and neural timing, or when delivered ventilator support does not match the patient’s respiratory effort.
3Mechanical breath delivery can be analyzed in 3 variables: trigger, control variable for flow/pressure delivery, and cycling-off variable; this framework helps classify asynchrony.
4The main clinical task is to inspect the patient’s breathing pattern together with the ventilator screen, because waveform-only observation can miss asynchrony and bedside observation is emphasized.
5Clinical signs that may suggest asynchrony include agitation, diaphoresis, tachycardia, hypertension, and failure of weaning or NIV.
6Before attributing problems to patient–ventilator interaction, airway complications such as partial obstruction or displacement, and major clinical changes such as pneumothorax or acute pulmonary oedema, should be excluded.
7Traditional detection used oesophageal and gastric pressure monitoring, but bedside recognition can be done less invasively using routinely displayed flow and pressure waveforms plus patient observation.
8Waveform abnormalities can be diagnosed with a stepwise six-item review of the pressure and flow traces: whether the breath was triggered, the early inspiratory pressure, the end-inspiratory pressure, the expiratory flow shape, the plateau pressure on an end-inspiratory hold, and auto-PEEP on an end-expiratory hold.
9The common types of asynchrony are grouped as triggering-phase asynchronies, pressure/flow-delivery asynchronies, and cycling-off asynchronies.
10Triggering-phase asynchronies include ineffective efforts, triggering delay, and autotriggering.
11Pressure/flow-delivery asynchronies include excessive ventilatory assist and insufficient ventilatory assist.
12Cycling-off asynchronies include delayed opening of the expiratory valve, premature opening of the expiratory valve, and double-triggering.
13Common asynchronies are triggering, cycling, and flow asynchrony, and in NIV they are trigger delay, ineffective inspiratory efforts, double triggering, auto-triggering, premature cycling, and delayed cycling.
14Ineffective effort is defined as an inspiratory effort not followed by ventilator triggering.
15Triggering time delay is the interval from onset of the triggering effort to onset of flow delivery.
16On the waveforms it appears as a lag between the start of the patient's neural inspiration and the start of the mechanical breath, which is linked to dynamic hyperinflation with flow still present at end-expiration.
17Ineffective triggering is described as one of the most frequent asynchronies and accounted for 88% of all asynchronous breaths.
18Ineffective triggering accounted for most asynchronous episodes, with autocycling next most common.
19In volume-assist ventilation, ineffective inspiratory efforts can be detected on the pressure waveform, whereas in pressure support they are best seen on the flow waveform; an abrupt increase in flow suggests an ineffective inspiratory effort.
20A waveform criterion for ineffective triggering is an inspiratory effort plus an abrupt airway pressure drop of 0.5 cmH2O with simultaneous inspiratory flow increase during inspiration, or a small expiratory flow and pressure decrease during expiration, without an assisted cycle.
21Clinically, wasted inspiratory efforts may be visible at the bedside.
22A trigger response time should be <100 ms.
23The effort needed just to trigger a breath contributes about 5% of total respiratory effort, and flow triggering lowers respiratory effort only slightly compared with pressure triggering; study figures were 17% with pressure triggering and 12% with flow triggering.
24Flow triggering is described as usually set by default at 2–5 L/min, with operator adjustment possible.
25Autotriggering is a ventilator-delivered cycle without a preceding patient inspiratory effort or airway pressure decrease.
26Double triggering is defined as 2 cycles separated by a very short expiratory time of <1/2 of the mean inspiratory time, together with an inspiratory effort.
27Reverse triggering is described as a pattern in deeply sedated patients in whom mechanical insufflation triggers a muscular effort, generating a “patient-triggered” breath.
28Flow asynchrony occurs when the patient’s flow demand does not match ventilator flow delivery.
29Flow starvation is described particularly in VC, where patients may show high ventilatory demand and a pressure–time waveform that is “sucked down”.
30Shorter rise time and higher flow cycle should be considered in tachypnoeic patients, whereas slower rise time may feel better in patients with low respiratory drive.
31Insufficient ventilator assist on pressure support is associated with a rounded or constant inspiratory flow waveform and vigorous inspiratory efforts causing a marked fall in Paw from the expected square shape.
32Cycling-off asynchrony occurs around the switch from inspiration to expiration.
33In premature cycling, the patient’s inspiratory effort continues into expiration, causing double-triggering and breath stacking.
34Delayed cycling refers to the opposite problem, with ventilator inflation continuing too long relative to the patient’s neural inspiration.
35If the breath lasts longer than the patient wishes, the patient may start to breathe out actively while the ventilator is still inflating, so airway pressure rises at the end of inspiration and expiratory flow increases.
36In pressure support, the expiratory trigger is based on a percentage of peak inspiratory flow, and most ventilators default to 25% of peak flow; the adjustable range is 5% to 90% in some descriptions, whereas it is 1% to 90%, or 5 to 25 L/min on some older ventilator software.
37A default expiratory trigger of 25% of peak flow is also given, with possible variation from 1% to 90% or 5 to 25 L/min in some older software.
38In ARDS, patients tend to synchronize better with late cycling-off such as 5% PIF, whereas COPD patients tend to synchronize better with early cycling-off such as 40–70% PIF.
3940%–70% decay of peak inspiratory flow improves synchrony in obstructive pulmonary disease, while 5% of peak inspiratory flow might be optimal in ARDS or acute lung injury.
40Premature cycling-off in an ARDS patient may show a reversed “tail-wing” expiratory flow pattern with concavity early in expiration.
41Severity can be quantified with the asynchrony index.
42AI is defined as the number of asynchrony events divided by the sum of ventilatory cycles.
43Another gives AI (%) = number of events/(ineffective breaths + ventilator cycles) × 100, and labels AI >10% as severe.
44One study found 24% of patients had AI >10%.
45When asynchrony accounts for >10% of total breaths it can increase sedation need and reduce sleep quality.
46Increased mortality is also associated with >30% asynchronous breaths.
47Reported frequency also varies by setting: overall patient-ventilator asynchrony is about 23%, although up to 93% of patients have at least 1 episode; significant asynchrony occurs in 25% of intubated patients, especially at the start of ventilation or after prolonged ventilation, and in nearly 50% with NIPSV; in NIV it is 24%–43%.