Management1Once mechanical ventilation is required, the next step is to choose the mode of ventilation, which specifies how the mechanical breath is delivered.
2Assist-control mechanical ventilation is described as the most widely used mode.
3In assisted-controlled ventilation, the ventilator provides most of the respiratory power and control while still allowing patient respiratory drive and muscle activity, which is preferable to total abolition of effort when possible because it may minimize respiratory muscle weakness and facilitate weaning.
4In assist-control ventilation, every patient-triggered breath is fully supported with the same control characteristics as mandatory breaths, using either a pressure target or a volume target.
5A/C can guarantee a minimum number of mandatory breaths; in AC-PC, the clinician sets the minimum mandatory breath number, inspiratory pressure, inspiratory time, PEEP, and FiO2, and the delivered tidal volume then depends on lung and chest wall compliance.
6SIMV guarantees a minimum number of fully assisted breaths that can be either pressure-targeted or volume-targeted, while allowing intermittent spontaneous breaths between them.
7For SIMV, the clinician sets the minimum number of mandatory or assisted breaths with either a volume or pressure target, pressure support for patient-triggered breaths, PEEP, and FiO2.
8Compared with A/C, SIMV differs because additional patient-generated breaths above the preset rate are not identical mandatory breaths; they are spontaneous or pressure-supported according to patient effort.
9One advantage attributed to SIMV in less sedated patients is attenuation of air trapping and hyperventilation that may occur with A/C.
10However, SIMV has drawbacks: increased work of breathing is identified as a major limitation, and SIMV has not been shown to be an effective weaning mode.
11PSV is used in awake patients who are assuming part of the work of breathing, and is commonly used for weaning.
12In PSV, the ventilator provides a preset positive pressure in response to inspiratory effort, holds that pressure during inspiration by gas flow from the ventilator, and cycles off when inspiratory flow falls below the set threshold.
13Because PSV has no set machine rate, minute ventilation can vary if respiratory drive is unreliable.
14PSV is the ventilator mode that most closely resembles spontaneous breathing and therefore promotes patient control and comfort.
15For pressure support settings during weaning, reduce PSV in steps of 2–4 cm H2O at least twice daily according to clinical tolerance.
16Extubation is generally considered after tolerance of about 8 cm H2O of PSV without PEEP, although this may vary with overall clinical status and endotracheal tube diameter.
17Pressure support is gradually reduced according to patient comfort and blood gases until the level is ~5 cmH2O, after which extubation can be considered if other criteria are met.
18During PSV weaning, respiratory rate should generally remain <30 breaths/min.
19For patients on IMV or SIMV during weaning, the mandatory rate can be gradually reduced while spontaneous breaths are commonly pressure-supported to overcome circuit and ventilator valve resistance.
20Daily trials of unassisted or spontaneous breathing are described as integral to evaluating readiness for separation from the ventilator.
21Daily sedation interruption and daily spontaneous breathing trials are identified as strategies that reduce duration of intubation or mechanical ventilation.
22Allowing patient-triggered breaths and avoiding unnecessary heavy sedation or neuromuscular paralysis may help preserve diaphragm strength and facilitate transition to spontaneous breathing.
23ASV is a closed-loop pressure-targeted mode that adjusts ventilator output breath by breath according to respiratory mechanics and breathing effort, with the goal of delivering a preset minute ventilation while minimizing work of breathing.
24Under standard conditions in ASV, 100% minute volume corresponds to 0.1 L/kg/min based on predicted body weight; thus a 70-kg patient receives 7.0 L/min at 100%, and 10.5 L/min at 150%.
25ASV adjusts inspiratory pressure to achieve a respiratory pattern of tidal volume and frequency that minimizes work of breathing, but the available tidal-volume range is constrained by the clinician-selected maximum pressure, Pmax.
26In ASV, breaths alternate between pressure-controlled and pressure-supported breaths depending on whether spontaneous breathing efforts are absent or present.
27During mandatory ASV breaths, inspiratory time and I:E are adjusted from the measured expiratory time constant; lower compliance shortens I:E, whereas higher resistance lengthens I:E to avoid air trapping.
28Pressure-regulated volume control and volume support are pressure-targeted modes in which the ventilator automatically adjusts inspiratory pressure to maintain a target volume; PRVC is time-cycled and volume support is flow-cycled.
29PRVC can operate as an A/C mode, where every patient-triggered breath is identical to the preset machine breaths, or as SIMV, where mandatory breaths are PRVC but patient-triggered breaths receive the preset pressure support instead.
30Proportional assist ventilation calculates patient mechanical properties using controlled test breaths, then offsets a clinician-selected percentage of the patient’s work of breathing by adding pressure, flow, and volume.
31NAVA uses an esophageal catheter with electromyographic electrodes at the diaphragm level; the electrical signal is used to trigger breaths, adjust delivered flow, and cycle ventilator breaths.
32These proportional modes are described as potentially improving patient–ventilator synchrony, and small trials show that they perform as designed, but randomized trials have not shown superior outcomes over more conventional approaches.
33Because NAVA and PAV are proportional modes, they depend on preserved respiratory drive, and choosing unsuitable patients for them can leave clinicians dissatisfied.
34Monitoring of EAdi separate from NAVA may also be useful for titrating other ventilation modes, though this requires further study.
35Routine monitoring of mechanically ventilated patients includes evaluation of respiratory system mechanics and waveform analysis.
36For general ventilator settings, tidal volume should be <=6 mL/kg PBW in ARDS, and 6–8 mL/kg PBW is an appropriate initial setting in most patients. A low tidal volume strategy of 4–8 mL/kg PBW in ARDS is also recommended, with tidal volume limited to 4–8 mL/kg PBW and plateau pressure kept below 30 cmH2O.
37An initial respiratory rate of 12–16 breaths/min is often used and then adjusted according to pH and PaCO2 to achieve the desired minute ventilation.
38Extrinsic PEEP is usually started at 5 cm H2O, and higher PEEP may be needed in acute hypoxemic respiratory failure with care to keep plateau pressure below 30 cm H2O; in severe respiratory failure, PEEP may need to be above 10 cm H2O, but it rarely needs to be above 20 cm H2O because of cardiorespiratory complications and alveolar overdistension.
39FiO2 is often initially set at 100% and then weaned rapidly to the minimum needed for adequate oxygenation; FiO2 <=60% is preferred to limit injury from prolonged higher oxygen exposure.
40In patients with dynamic hyperinflation or auto-PEEP, suggested corrective changes include improving airflow resistance, extending expiratory time, reducing minute ventilation, decreasing respiratory rate and tidal volume, or increasing inspiratory flow rate to allow more exhalation time.
41If hypotension develops with auto-PEEP, temporarily disconnect the patient from the ventilator until hypotension resolves, alongside volume resuscitation and ventilator adjustment.
42Management of ineffective efforts includes boosting respiratory drive by reducing assist, lightening sedation, or correcting alkalaemia, reducing dynamic hyperinflation by lowering airway resistance, lowering assist, or raising the cycling-off flow threshold, and, in patients with dynamic hyperinflation, adding external PEEP to offset intrinsic PEEP.
43Double triggering can be minimized by decreasing trigger sensitivity or increasing inspiratory time.
44If a patient continues to generate significant inspiratory efforts after ventilator and other causes are addressed, judicious sedation is advised.