Management1Mechanical ventilation may be noninvasive via mask or invasive via endotracheal tube, and invasive respiratory support requires tracheal intubation.
2The decisions to provide mechanical support and to perform tracheal intubation should be made independently, particularly because noninvasive options such as CPAP can support ventilation.
3NPPV is usually not the definitive treatment for pneumonia or ARDS unless the patient is plainly improving or has do-not-intubate status; otherwise, endotracheal intubation with mechanical ventilation is preferred.
4Ventilator modes differ by how breaths are triggered, targeted, and cycled.
5In all modes, a breath is triggered either by a timer at a set respiratory rate or by patient effort.
6A breath may terminate by delivery of a set tidal volume, completion of a set inspiratory time, or decrease in flow to a predetermined percentage of peak flow.
7Assist-control mode supports breaths at a minimum set respiratory rate and is described as the most widely used mode.
8In assist-control mechanical ventilation, each assist or control breath delivers a preset tidal volume or pressure, is either patient-triggered or time-triggered, and inspiration ends either by delivery of preset tidal volume in volume-controlled breaths or by time in pressure-controlled breaths.
9Alveolar hyperventilation may be a hazard in assist-control ventilation.
10SIMV provides mandatory breaths at a preset rate while allowing spontaneous breathing between mandatory breaths.
11In SIMV, mandatory breaths may be time-triggered or patient-triggered, and there is a synchronization window in which the ventilator responds to patient effort.
12The major drawback of SIMV is increased work of breathing.
13Pressure support ventilation provides a driving pressure consisting of inspiratory pressure and PEEP to support patient-initiated breaths.
14PSV is patient-triggered and flow-cycled, and inspiration ends when inspiratory flow falls below either <5 L/min or <25% of peak inspiratory flow depending on ventilator type.
15PSV has no set machine rate and is commonly used for weaning.
16In assist-control pressure control, the clinician sets the minimum number of mandatory breaths, inspiratory pressure, inspiratory time, PEEP and FiO2. The delivered tidal volume depends on lung and chest wall compliance. When pressure control is time-cycled, a pre-set pressure is applied throughout inspiration and VT depends on compliance and airway resistance.
17For volume-control ventilation, the set variables are tidal volume, PEEP, RR, and flow pattern.
18For initial adult invasive settings, start with a tidal volume of 8 mL/kg predicted body weight and, if possible, reduce it to 6 mL/kg over the next 2 hours; patients with ARDS should receive tidal volume ≤6 mL/kg predicted or ideal body weight.
19An initial tidal volume of 6–8 mL/kg ideal body weight, rate 12–20 breaths/min, inspiratory time 0.7–1.2 s, inspiratory flow about four times minute ventilation, PEEP 4–5 cm H2O, FiO2 1.0, and plateau pressure <30 cm H2O is recommended; the lung-protective protocol instead starts at 8 mL/kg PBW and lowers it by 1 mL/kg every 2 hours to 6 mL/kg, with PEEP 5 cm H2O and the lowest FiO2 that achieves an SpO2 of 88–95%.
20Typical adult IPPV starting values are VT ~6–8 mL/kg, respiratory frequency ~8–14 breaths/minute, minute ventilation ~6 L/min, PEEP ≥5 cmH2O, and I:E ratio ~1:2.
21It is suggested to select assist-control as the initial mode and an initial tidal volume of 8 mL/kg predicted body weight, then reduce to 6 mL/kg over the next 2 hours if possible.
22For respiratory rate, 12–16 breaths/min is often chosen initially and then adjusted according to pH and PaCO2.
23Respiratory rate should be set based on indication, including 20/min for sepsis, metabolic acidosis, or high minute ventilation requirements, even higher in ARDS with low tidal volumes, and 10–12/min in obstructive disease to allow exhalation time.
24Extrinsic PEEP is typically set at 5 cm H2O to prevent end-expiratory alveolar collapse or atelectasis.
25PEEP starts at 5 cmH2O, but in ARDS it is commonly titrated, for example to a PaO2/FiO2 target such as the ARDS Network protocol, and its effect depends on how recruitable the lung is, the duration of ARDS, and phenotype, because PEEP may cause alveolar recruitment, overinflation, or both.
26Starting with a higher PEEP in ARDS is advised, such as 10–14 mbar or an ARDSNet-style FiO2/PEEP table from FiO2 0.3 with PEEP 5 up to FiO2 1.0 with PEEP 20–24, but recruitability varies widely and ARDS duration (late ARDS is less recruitable) and phenotype decide whether PEEP recruits lung, overinflates it, or both.
27When an obstructed patient is breathing in assisted modes, external PEEP at about 80% of PEEPi can reduce inspiratory elastic work at triggering without worsening dynamic hyperinflation; however, during initial controlled ventilation with excessive PEEPi and dynamic hyperinflation, when work of breathing is not an issue, extrinsic PEEP increases lung volumes and should not be routinely used because benefit is uncertain and responses are unpredictable.
28FiO2 is often initially set at 100% and then quickly weaned to the minimum needed for adequate oxygenation.
29FiO2 ≤60% is preferred to reduce injury from prolonged exposure to higher oxygen concentrations.
30Once stabilized, FiO2 should be titrated to maintain PaO2 >60 mmHg or SpO2 >93–94% in a normal lung and 88–92% in hypercapnic respiratory failure, whereas the aim is SaO2 94–98% in patients at low risk of hypercapnic respiratory failure (92–98% if over 70 years old) and 88–92% pending ABG analysis in those at risk.
31FiO2 should be titrated down from 100% to maintain oxygen saturation of 90% at the lowest possible FiO2.
32Oxygenation is mainly determined by FiO2, PEEP, and mean airway pressure, while PaCO2 is determined by minute ventilation.
33If the goal is to change PaO2 or SpO2, adjust FiO2 and PEEP, whereas if the goal is to increase pH, increase ventilation parameters.
34Routine ventilator monitoring includes assessment of respiratory mechanics and waveform analysis.
35During constant-flow volume-control ventilation, an end-inspiratory hold allows measurement of plateau pressure.
36Static compliance is calculated as VT/(Pplat − PEEP).
37In ARDS, the targets are Pplat <30 cmH2O and driving pressure <14 cmH2O.
38One table lists plateau pressure <30, driving pressure (Pplat − PEEP) <15, compliance >50 and a PIP − Pplat value labelled airway resistance of <10; it prints no units. A PIP − Pplat difference is a pressure gradient, and resistance itself is that gradient divided by inspiratory flow. Another view treats a PIP − Pplat difference above 5 cm H2O as typically representing increased airway resistance. The two PIP − Pplat figures differ (<10 without a unit, and above 5 cm H2O), so no single cut-off is stated here.
39If auto-PEEP or breath stacking is detected, improving airflow resistance, extending expiratory time, and reducing minute ventilation help reverse the problem.
40Decrease respiratory rate or increase inspiratory flow to increase exhalation time when auto-PEEP is present.
41If severe auto-PEEP causes hypotension or obstructive shock, temporarily disconnect the ventilator circuit to allow complete exhalation.
42If hypotension occurs from obstructive lung disease with auto-PEEP, patients should be temporarily disconnected from the ventilator until hypotension resolves and volume resuscitated in addition to ventilator adjustment.
43If sudden difficulty occurs with high pressure during volume-preset ventilation or with oxygenation, an immediate search is advised, beginning with airway patency, then the circuit and ventilator, and then factors affecting elastance and resistance such as bronchospasm, secretions, pneumothorax, and asynchronous breathing.
44Ventilator alarms for pressure, volume, and temperature should be set appropriately, and may be temporarily but never permanently disabled while troubleshooting.
45Mechanical ventilation goals include optimizing oxygenation and ventilation, offloading work of breathing, and protecting the airway.
46Analgesia and sedation are required for comfort and ventilator synchrony, and should be titrated to comfort and therapeutic goals while avoiding over- and under-sedation.
47Deep sedation is associated with delirium, prolonged mechanical ventilation, and mortality, so lighter sedation is reasonable for most patients if comfort is maintained.
48Deep sedation is required after recent neuromuscular blockade or during procedures or diagnostic studies.
49The RASS is described as ranging from +4 to −5, and sedation should be maintained at the highest RASS score at which the patient is comfortable.
50An ABCDEF bundle includes pain assessment and management, daily sedative interruption and spontaneous breathing trials, optimal sedation strategy, delirium prevention and assessment, early mobilization, and family engagement.
51Measures to reduce VAP risk include elevating the head of the bed to 30 to 45 degrees.
52For weaning, daily assessment of readiness and minimization of sedative use are recommended.
53Objective readiness criteria include improvement in the cause of respiratory failure, adequate oxygenation, arterial pH >7.25, hemodynamic stability, and ability to initiate a breath.
54A daily screen using RSBI is recommended, with RSBI <105 allowing progression to a spontaneous breathing trial and RSBI >105 arguing against a weaning trial that day.
55A similar but not identical threshold is an RSBI value <100, which makes success more likely.
56Candidate criteria for weaning are PaO2/FiO2 >150–200, or 120 in chronic hypoxemia, or SpO2 >90% on FiO2 <40% and PEEP 5–8 cmH2O, with PEEP 8 cmH2O accepted in obese patients, abdominal distention, or endotracheal tube <7.0 mm, and FiO2 50% accepted in significant underlying lung disease.
57Spontaneous breathing trials may be performed with a T-piece or minimal support such as low pressure support, automatic tube compensation, or CPAP.
58An SBT of 30 minutes is recommended, extended to 120 minutes after prolonged ventilation >10 days; patients who tolerate an SBT of at least 30 minutes and no more than 2 hours without signs of distress are extubated.
59SBTs or weaning support over 30–120 min include CPAP with PEEP 5 cmH2O and pressure support 5–8 cmH2O.
60Gradual reduction of pressure support is described as the preferred weaning technique.
61SIMV weaning is described as progressive reduction in mandatory breath frequency, with spontaneous breaths usually pressure-supported.
62T-piece trials and PSV gave better results than SIMV in patients with complex respiratory problems.
63CPAP can be used during weaning with SIMV or pressure support, during spontaneous breathing trials, and before extubation to help prevent alveolar collapse, hypoxaemia, and loss of compliance.
64Noninvasive respiratory support such as BiPAP or CPAP can be used after extubation to prevent respiratory failure and re-intubation.