Management1ICDs are placed for either primary prevention of SCD or secondary prevention after cardiac arrest, and survival with ICDs is better than with antiarrhythmic therapy in selected at-risk patients.
2ICD implantation should be considered in patients on optimal medical therapy who have estimated survival of at least 1 year with good functional status.
3Expected survival with good functional status of at least 1 year is similarly specified as part of the indication summary.
4For secondary prevention, survivors of SCD without a reversible cause derive major benefit from ICD implantation in preventing death from VT or VF.
5A Class I secondary-prevention indication is ischemic heart disease with sudden cardiac arrest due to VF or VT, or stable sustained VT, not caused by a transient or reversible event, with meaningful survival expected to be >1 year.
6Ischemic heart disease with unexplained syncope plus induced sustained monomorphic VT during electrophysiologic study is also included as a Class I indication when meaningful survival is expected to be >1 year.
7Secondary-prevention Class I indication is survival after cardiac arrest due to VF or sustained VT without reversible causes.
8Indications include documented cardiac arrest due to VF not due to a reversible cause, and documented sustained VT either spontaneous or induced during EPS not associated with acute MI or a reversible cause.
9For primary prevention after myocardial infarction, patients on guideline-directed medical therapy whose infarct was at least 40 days earlier and whose left ventricular ejection fraction is ≤30% are at high risk of sudden cardiac death, and ICD implantation is given in ischemic cardiomyopathy when EF ≤35%, with the MI at least 40 days or revascularization at least 3 months earlier.
10Prior MI with EF ≤30% meets MADIT II criteria, and prior MI with EF ≤35% plus inducible sustained VT or VF on EPS when MI and EPS are each >4 weeks from the infarct meets MADIT I criteria.
11SCD-HeFT criteria are ischemic dilated cardiomyopathy with documented prior MI, NYHA class II or III heart failure, and EF ≤35%, provided at least 40 days have passed since the most recent MI, at least 3 months since CABG or PCI, medical therapy has been optimized, and expected survival with good functional status is at least 1 year.
12Class I indications are depressed LV function at least 40 days post-MI and at least 90 days post-revascularization with persistent heart failure symptoms, expected survival >1 year, as well as nonsustained VT due to prior MI with depressed left-sided systolic function and inducible VT or VF during EPS.
13Primary-prevention indications are ischemic cardiomyopathy with LVEF ≤30%–40% at least 40 days post-MI or at least 3 months post-CABG, LVEF ≤30% with NYHA class I, II, or III, LVEF ≤35% with NYHA class II or III, LVEF ≤40% with inducible VF or sustained VT at EP study, and syncope of unknown origin with VT/VF induced at EP study.
14That same source also lists spontaneous sustained VT, with or without hemodynamic instability, in structural heart disease as a Class I indication.
15For nonischemic cardiomyopathy, depressed LV function is a Class I indication when meaningful survival is expected to be >1 year.
16Nonischemic cardiomyopathy with at least 3 months of documented CHF, LVEF ≤35%, and NYHA class II or III is specified as a primary-prevention indication.
17In DEFINITE, patients with nonischemic cardiomyopathy, EF <35%, and nonsustained VT had a strong but not statistically significant trend toward lower all-cause mortality with ICDs.
18Patients meeting SCD-HeFT criteria—ischemic or nonischemic cardiomyopathy, EF ≤35%, and NYHA class II or III heart failure—qualify for an ICD, provided at least 40 days have passed since the most recent MI, at least 3 months since revascularization with CABG or PCI, medical therapy has been optimized, and expected survival with good functional status is at least 1 year.
19Additional Class I groups include sudden cardiac arrest or sustained ventricular arrhythmias due to nonischemic cardiomyopathy, ARVD, HCM, cardiac sarcoidosis, neuromuscular disorders, cardiac channelopathies, symptomatic long QT syndrome failing beta-blocker therapy, catecholaminergic polymorphic VT failing beta-blocker therapy, Brugada syndrome, early repolarization, short QT syndrome, idiopathic polymorphic VT, and congenital heart disease.
20ICD benefit is clear in survivors of SCD and in patients with documented VT/VF outside the early post-infarct phase who have non-inducible VT/VF at EPS, inducible VT/VF resistant to treatment, or VT/VF with LVEF ≤30% regardless of EPS-directed drug results.
21Modern ICDs are typically implanted percutaneously with transvenous leads rather than epicardial leads requiring thoracotomy.
22A typical ICD consists of a power source, electronic circuitry, and a lead system; the power source is lithium-based and battery life is 5–10 years, largely determined by shock frequency.
23All ICDs are also ventricular pacemakers, and some patients with depressed LVEF and wide QRS may receive a CRT-D that combines biventricular pacing with shock capability.
24If dual-chamber pacing is needed, a second lead is placed in the right atrium.
25ICD leads differ from standard pacemaker leads by having defibrillation coils; at least one RV coil is required for defibrillation, and some models have a second coil in the superior vena cava/right atrium.
26The surface of the ICD itself functions as one of the electrodes in the defibrillation system.
27ICDs can detect and terminate life-threatening ventricular tachyarrhythmias using overdrive pacing, cardioversion, or defibrillation.
28Modern ICDs may have leads to sense and pace both the right atrium and ventricle, and the lithium batteries can provide energy for >100 shocks of about 30 J each.
29ICD discharges are painful if the patient is conscious, but VT may often be terminated painlessly by overdrive pacing.
30The ICD’s diagnostic and treatment functions are set at the time of implantation, when cardioversion and defibrillation thresholds are determined by inducing VT and VF and adjusting the shock strength to a level above the minimum needed to end the induced rhythm. For subcutaneous ICDs, the HRS/EHRA/APHRS/SOLAECE consensus recommended defibrillation efficacy testing, at a time when no data addressed the safety and efficacy of omitting it with that device. The newer PRAETORIAN-DFT randomized trial, a multinational study in subcutaneous ICD patients from 37 centers, enrolled 965 patients followed for a median of 41 months. In the no-testing arm the PRAETORIAN score was evaluated before discharge, and omission was conditional: with a score of 90 or higher, the protocol told physicians to do at least defibrillation testing. In that arm 459 of 474 patients had a low score and 15 (about 3%) an intermediate or high one; testing was attempted in 14 of those 15. The score is a chest X-ray based predictor of defibrillation outcome. The primary end point was a failed first shock for spontaneous ventricular arrhythmia, tested for noninferiority against a 3% absolute margin; it occurred in 1.7% without testing versus 2.3% with testing, and the authors concluded that score-guided omission did not raise that risk and reduced procedural risk without more revisions. For transvenous devices, Omission reasonable in an initial left-pectoral implant when sensing, pacing and impedance are appropriate with a fluoroscopically well-positioned RV lead; testing is reasonable for right-pectoral implants or generator changes. Two randomized trials in first-time transvenous ICD recipients agree with omission: SIMPLE enrolled patients receiving a first ICD for standard indications, excluded those expected to need a right-sided implant, and found that routine testing neither improved shock efficacy nor lowered arrhythmic death; NORDIC ICD found shock efficacy not inferior without testing in 40 J devices and concluded that testing should no longer be recommended for routine left-sided first implants. At transvenous ICD implantation testing should not be performed in patients with a documented nonchronic cardiac thrombus, or AF or flutter without adequate anticoagulation; the same sentence goes on to name critical aortic stenosis, unstable CAD, recent stroke or TIA and haemodynamic instability, and ends with other known morbidities associated with poor outcomes, so these named conditions are examples, not a complete list.
31VT is typically treated by low-energy shocks or antitachycardia pacing, and in VF the ICD can deliver up to 5 additional shocks if the first shock fails.
32ICD programming uses up to 4 ventricular tachyarrhythmia zones, including a monitor-only zone, optional slow and fast VT zones, and a VF zone.
33The different VT/VF zones are programmed with ATP and shocks, usually 35–42 J, and ATP is painless whereas shocks are painful and reduce battery life by about 1 month per full-energy shock.
34Algorithms such as morphology recognition are used to distinguish SVT from ventricular arrhythmias so that therapy can be withheld for SVT.
35Favorable ICD outcomes are reported compared with drug therapy: sudden death and total mortality of 3.5% and 13.6% with ICD, versus 12% and 34% with empirical amiodarone, 14% and 24% with EPS-directed drug treatment, and 3.7% and 37% with surgery.
36There is a 31% reduction in 3-year mortality versus antiarrhythmic drug therapy in SCD survivors, survival benefit persisting for at least 8 years, and improved survival in patients with HCM.
37ICD use has reduced the sudden death rate in patients with a history of serious ventricular arrhythmias to about 2% per year.
38CASH showed a 23% mortality reduction with ICD versus metoprolol or amiodarone, and AVID showed mortality reductions of 39%, 27%, and 31% at 1, 2, and 3 years versus antiarrhythmic drug therapy.
39AVID showed over 50% reduction in arrhythmic death, with the greatest survival gain in patients with EF 20%–35%.
40In primary prevention, there is a 55% reduction in total mortality in MADIT I, a 29% reduction in all-cause mortality in MADIT II, and a 23% reduction in overall mortality in SCD-HeFT.
41Trial results disagree: DINAMIT found that ICD reduced arrhythmic death but did not reduce overall mortality in patients implanted 6–40 days post-MI, and CABG Patch found no significant improvement in overall mortality when ICD was implanted at the time of bypass surgery, although arrhythmia-associated death fell relatively by 45%.
42Current practice often combines an ICD with low-dose amiodarone to improve arrhythmia control, reduce atrial tachyarrhythmias, slow VT rate, and prolong battery life by reducing arrhythmia frequency.
43Antiarrhythmic drug-only strategies are becoming less common as ICD availability widens.
44For a patient reporting a single ICD shock, assessment should include acute cardiac ischemia, worsening chronic CHF, symptoms of new-onset heart failure, and electrolyte abnormalities.
45If multiple ICD discharges are confirmed on interrogation, emergent consultation is required, and reprogramming may be needed based on interrogation findings.
46If a lead problem is identified, reimplantation is required.
47A magnet can be placed over the ICD to disable defibrillator therapies, including ATP and defibrillation, if the clinician is confident the shocks are inappropriate, such as shocks delivered for supraventricular tachycardia.
48If frequent ventricular ectopy is present after inappropriate or repeated shocks, amiodarone 150 mg IV/IO bolus followed by 1 mg/min for 6 hours and then 0.5 mg/min for 18 hours is recommended; if unavailable or ineffective, lidocaine 1–1.5 mg/kg IV/IO bolus followed by infusion 1–4 mg/min can be considered.
49In patients with an ICD who have a history of serious ventricular arrhythmias, the device lowers the sudden-death rate to about 2% per year.
50During resuscitation of a patient with an ICD, shockable rhythms should be treated with standard transthoracic defibrillation using monophasic or biphasic defibrillators, and the defibrillation pads should be placed at least 8 cm from the device, using anteroposterior positions if possible for a device implanted below the right clavicle.
51If VT/VF arrest persists after device failure is presumed, 300 mg amiodarone IV push is recommended, followed by an additional 150 mg IV/IO if VT/VF persists, and lidocaine 1–1.5 mg/kg IV/IO if amiodarone is unavailable or ineffective.