Recognition and diagnosis1TCD is used to study intracranial arterial circulation by insonating basal cerebral vessels through acoustic windows and identifying vessels by depth and flow direction; TCCD adds B-mode and color imaging to visualize the vessel and optimize the Doppler angle.
2There are four acoustic windows described for neurocritical care TCD: transtemporal, suboccipital/transforaminal, submandibular, and transorbital.
3The transtemporal window is located above the zygomatic arch between the tragus and the lateral eye canthus, and it is the most commonly used window for the MCA, ACA, terminal ICA, and PCA.
4The suboccipital/transforaminal window passes through the gap between the occipital bone and the atlas vertebra or through the foramen magnum, and it is used to insonate the vertebral arteries and basilar artery.
5The submandibular window is obtained through the submandibular area near the angle of the mandible and is used to insonate the extracranial ICA; it is also used when standard windows are not possible.
6The transorbital window is obtained through the orbital plate and optic foramina/superior orbital fissure and allows insonation of the ophthalmic artery and the carotid siphon.
7For vessel localization, typical insonation depths from the transtemporal window are MCA 40–60 mm and terminal ICA 55–65 mm, ACA 60–70 mm and PCA 65–70 mm, and from the suboccipital window vertebral artery 50–70 mm and basilar artery 80–100 mm.
8Transtemporal targeting depths and angulations are ACA with anterior angulation at 6–7 cm, terminal ICA with caudal angulation at 6–7 cm, and PCA with posterior angulation at 5.5–7.5 cm.
9For MCA interrogation, the MCA is insonated through the transtemporal window at a depth of 40–60 mm, and the time to obtain an adequate MCA echographic image is estimated at 50 ± 20 seconds.
10Bilateral pulsed-wave Doppler tracings should be recorded for at least 10 cardiac cycles after a 30 seconds stabilization period, and the protocol is then repeated on the opposite hemisphere.
11The Doppler principle in TCD is based on ultrasound crossing the skull and reflecting from erythrocytes, producing a Doppler shift directly proportional to erythrocyte velocity.
12Intracranial Doppler signals normally show a prominent diastolic component.
13The angle of insonation is a major determinant of Doppler accuracy. The need to align the beam with the direction of blood flow is the major limitation of Doppler, and the maximum angle for accuracy is 15°. For transcranial midline-shift (MLS) measurement, the guideline upward angle of insonation is < 10–15°, though this may not always be possible. No data link the angle of insonation to the accuracy of transcranial MLS measurements.
14Color coding in TCCD is described as blue for flow away from the probe and red for flow toward the probe.
15TCD is used to monitor for arterial cerebral vasospasm in aneurysmal SAH, and this is recommended with class IIa, level B; however, against angiography, TCD is reliable for detecting MCA and basilar artery vasospasm, whereas its sensitivity is lower for ACA and PCA vasospasm.
16Professional societies including the AHA, American Society of Neuroimaging, and AIUM are described as recommending TCD for vasospasm monitoring.
17For vasospasm diagnosis, MCA performance versus CT angiography is relatively good, with sensitivity/specificity of 89–98%, whereas ACA, PCA, basilar artery, and vertebral artery are less sensitive and specific than MCA, particularly through the transforaminal window.
18TCD also allows visualization of microembolic signals from intracranial or central sources and can stratify recurrent stroke risk in patients with embolic phenomena.
19TCD can diagnose intra- or extracardiac shunts and may help explain neurologic deterioration in systemic illness.
20TCD is described as an adjunct in assessment of brain death.
21Use is also cited for raised intracranial pressure and intracranial circulatory arrest, and TCD is cited as a noninvasive surrogate for ICP and CPP in some settings.