Management1Extracorporeal life support cannulation places two cannulas, either through a surgical cut-down or percutaneously by the Seldinger technique with serial dilations. A surgical cut down gives a direct view of the vessels.
2The percutaneous approach is preferred in adult and pediatric patients >10 kg, can be performed at the bedside, avoids vessel ligation, and reduces bleeding complications.
3Percutaneous cannulation may also be combined with surgical cutdown by exposing the vessel first and then using the Seldinger technique.
4Percutaneous insertion is described as ECHO-guided and preferred in emergency settings or when cannulation occurs outside the operating room.
5In difficult percutaneous access, vague pulses such as during CPR, or high-risk morbid obesity, venous cutdown may be used.
6Surgical cutdown can also be used when percutaneous cannulation is difficult or body habitus is unfavorable, and may be performed at the bedside or in the operating room.
7The open approach may be open cutdown alone, open cutdown plus Seldinger access, or open cutdown with an end-to-side Dacron graft to the artery for VA cannulation.
8A ligation cannulation technique should be used for an artery only when collateral flow is adequate or when a distal perfusion cannula is added; this means it may be useful in the femoral artery, but tying off the subclavian artery beyond the thyrocervical and thoraco-acromial branches will cause arm ischemia.
9Peripheral cannulation can be percutaneous, surgical directly into the vessel, or through a Gore-Tex side or chimney graft.
10Cannulation site selection depends on the kind of ECMO support required, the proceduralist’s preference, and the patient’s condition. For cannulation during cardiopulmonary resuscitation, chest compressions should be interrupted as little as possible throughout cannulation, which vessels are targeted depends on vessel size and on the operator(s), and the cannulation may be unilateral or bilateral.
11Femoral-femoral access offers easier access in emergencies, femoral-axillary can be chosen if femoral access is not possible, and central cannulation is commonly done in the operating room after cardiopulmonary bypass.
12Central cannulation via median sternotomy uses the RA for venous drainage and the ascending aorta for arterial return.
13Central cannulation is generally first choice for postcardiotomy shock and can be used as an upgrade from peripheral ECMO when peripheral flows are inadequate or when there is LV distention or differential hypoxia affecting the heart or brain.
14Central cannulation may also be considered for high-flow ECLS settings such as septic shock, absent or complicated peripheral vascular access, limb ischemia, need for additional devices, or vascular injury after peripheral cannulation.
15Real-time ultrasound-guided vascular puncture is recommended for faster access and to minimize vascular complications, and ultrasound is highly recommended to avoid multiple punctures and vascular complications.
16Ultrasound helps vessel sizing, localization, and confirmation of position; fluoroscopy helps guidewire insertion and cannula placement; chest X-ray can assess final cannula position.
17If ultrasound guidance is not available, anatomical landmarks may be used and venous placement can be confirmed with a transducer.
18TTE and TEE can both guide cannulation, but in mechanically ventilated adults TEE enhances guidewire visualization compared with TTE.
19Key TTE views are the subcostal IVC and descending aorta for VA ECMO, with subcostal RV inflow-outflow and parasternal RV inflow views helping visualize wires in the RA.
20Key TEE views are the midesophageal bicaval and IVC views, plus the descending aorta for VA ECMO.
21For successful cannulation, key technical steps are to identify the target vessels, use real-time imaging, and, for femoral arterial access, puncture the common femoral artery above its bifurcation; confirm venous and arterial guidewire position with ultrasound, TTE/TEE, or fluoroscopy, check that the wire tip moves freely, and direct the venous cannula to the mid right atrium.
22During VV echocardiographic guidance, the final cannula position should maximize drainage while considering the distance between cannulas, the direction of the reinjection jet, and the location of the main draining point in multistage cannulas.
23Because lung-rest ventilation may shift the diaphragm upward, the drainage cannula may be placed in the RA initially and reassessed after ventilator adjustment.
24The reinfusion cannula may not always be visualized in the SVC, but it should lie a few centimeters above the SVC-RA junction.
25For RA-PA cannulation, the wire should be visualized up to the PA, the cannula followed carefully over the wire, and the tip positioned 2–3 cm above the pulmonary valve without being selective.
26Cannula size selection aims to provide adequate flow with limited pressure drop, and vessel diameter can be estimated by ultrasonography with calculated circumference = π × D, where π = 3.14.
27The largest cannula with the minimum length is recommended to maximize flow, reflecting the relation Flow α Radius4 × pressure/Length.
28The Hagen-Poiseuille relationship is ∆P = 8µLQ/πR4, and length, and especially radius, strongly influence resistance and achievable blood flow; this relationship describes steady (laminar) flow in a rigid tube or catheter.
29Cannula size is usually smaller than the vessel maximum so some blood can pass around the cannula, reducing edema, stasis, venous hypertension, DVT, ischemic injury, and limb ischemia.
30One ultrasound-based sizing rule states Fr = 3 × narrowest vascular diameter in mm, with a small gap of a few Fr recommended.
31The drainage cannula should not exceed 2/3 of the total femoral vein diameter.
32For adults, ECMO access must be able to support blood flow equal to the patient’s full cardiac output, cited as 5 L/min, whereas no single circuit flow counts as full flow and flows should be goal-directed toward rapid lactate normalization, SvO2 >65-70%, and an age-appropriate mean arterial pressure, which in sepsis may need >150-200 ml/kg/min in young children.
33A suggested drainage cannula is 23–25 Fr and reinfusion cannula 17–21 Fr, with multistage drainage cannulas preferred and reinforced heparin/bioline-coated cannulas recommended.
34Venous cannulas are usually 15–29 Fr and arterial cannulas 15–23 Fr.
35Drainage cannulas are larger and typically 23–29 Fr.
36Generally 25–30 Fr multistage cannulas are used for venous drainage and 15 Fr, 17 Fr, or 19 Fr cannulas are used as supply cannulas.
37Avalon bicaval dual-lumen cannulas are reported in sizes 13–31 Fr.
38In pediatric patients, arterial / venous cannula sizes by weight (kg) are 2: 8 Fr / 8–10 Fr; 3–6: 10 Fr / 10–12 Fr; 6–8: 12 Fr / 14 Fr; 8–16: 14 Fr / 17 Fr; 16–30: 17 Fr / 19 Fr; 30–40: 17 Fr / 21 Fr; and >40: 21 Fr / 25 Fr; cannulas chosen from weight-based size charts usually deliver about 100 mL/kg/min for V-A flow and 80 mL/kg/min for (dl)V-V, but if 200 mL/kg/min is needed in univentricular or septic shock, the venous cannula can be upsized to double the bodyweight or two venous cannulas used, the arterial cannula upsized to 1.5 times the bodyweight while accepting a higher line pressure, and transthoracic cannulation considered if peripheral vessels will not accept cannulas that large.
39BioMedicus cannula sizes are usually 17F, 19F, or 21F NextGen BioMedicus.
40For central cannulation, venous options include CB Medtronic 35-cm DLP 32F, 36F, or 40F malleable venous; CB Medtronic DLP 2 stage 34/46; or DII 40F RA cannulas, with patient inflow using Carmeda EOPA 22F or 24F for PA or aortic return.
41General pressure-drop guidance to limit hemolysis is under 100 mmHg for the arterial return cannula and under 50 mmHg for the venous drainage cannula; however, manufacturer pressure-drop charts were historically generated using water, so they underestimate the cannula’s true resistance with blood and should be interpreted with blood’s higher viscosity in mind.
42If resistance to cannula insertion is excessive, it is advised to drop down one size and have one size above and below the expected size available.
43For limb perfusion during femoro-femoral VA ECMO, a 6 Fr distal perfusion catheter introduced a few centimeters downstream in the superficial femoral artery and connected to the Luer connector of the outflow cannula is recommended.
44Distal limb reperfusion uses a sheath in the superficial femoral artery in antegrade fashion, connected with a 3/8 × 3/8-inch connector with a Luer-Lok between the arterial cannula and pump tubing; the distal perfusion cannula is usually placed through the superficial femoral artery and sometimes through the posterior tibial artery in a retrograde fashion.
45If low flow or reduced heparinization is a concern in that distal perfusion system, heparin 0.5–2 U/kg/hr may be infused by pressure pump into the system.
46A neonatal semi-Seldinger approach for right atrial double-lumen VV ECMO via the right jugular vein uses a micropuncture needle, then a 5 Fr micropuncture sheath, followed by heparin administration, then a .025 inch guidewire advanced 10–12 cm until ectopics appear or resistance is felt, after which the wire is withdrawn slightly.
47In that neonatal technique, the cannula tip should end in the mid-right atrium at a depth of 6–8 cm.
48In older patients, VV cannulation can use a pure percutaneous technique.
49Patient preparation for bedside or ICU cannulation includes sterile positioning and draping, operating light, electrocautery, adequate room space, protection of lines and tubes, and preparation of a backup site in case of failure; for femoral access, both groins are usually prepared.
50In one approach, alcohol antiseptic at cannulation is advised against, to avoid an electric arc between the electrocautery and alcohol vapor. The same approach says the resuscitator bag should not be placed under or near the draped patient, because of oxygen-related surgical fire risk. In another approach, a chlorhexidine-alcohol skin antiseptic should be applied to the access site and allowed to dry before draping. One chlorhexidine-alcohol applicator label says not to drape the patient or bring in an ignition source such as cautery or laser until the solution is fully dry, which takes at least 3 minutes on hairless skin and up to 1 hour in hair. Another applicator label says the product should not be used with electrocautery procedures.
51Patients are often heavily sedated before cannulation, but teams may also cannulate awake, extubated patients with local anesthetic.
52After insertion, cannulas are secured to prevent dislodgement, and some operators support the tubing with a blanket or towel bundle to decrease the risk of dislodgement.
53Connections to the ECMO circuit should be made wet-to-wet to prevent leaks and air entrainment.
54Before starting flow, all air should be purged from circuit-cannula connections to reduce embolization risk.
55After connection, pump RPMs or blood flow are gradually increased: Blood flow is increased to 3–6 L/min or until VV arterial oxygen saturation is at least 80% or VA arterial pressure reaches target; VV flow is slowly increased to achieve target blood flow and arterial oxygen saturation >80%.
56With VA initiation, clamps are removed to establish antegrade ECMO flow, and mechanical chest compressions can then be stopped.
57A heparin bolus may be given once the system is connected and then adjusted per institutional protocol, although it may be omitted in hypothermic or near-drowning patients because of hemorrhage risk.
58In pediatric ECMO, an initial unfractionated heparin bolus of 50–100 units/kg is given at cannulation depending on preexisting bleeding, followed by UFH infusion 5–20 units/kg/hr once ACT drops to 300 unless there is excessive bleeding.
59For VA ECMO anticoagulation, an aPTT of 60–80 seconds is suggested to prevent circuit thrombosis, or 40–60 seconds if bleeding risk is higher; with lower anticoagulation targets, circuit flow should be kept >1 L.
60For patients with life-threatening bleeding, devices may be run without heparin for extended periods if pump stoppage and stagnation are avoided.
61In VV ECMO, hypotension during initiation may occur from a washout effect, so inotropic support may need to be increased and circuit calcium and potassium kept normal.
62After cannulation and through the run, care includes reassessing cannula position with echocardiography or radiography, checking secure fixation, examining for infection, bleeding, or hematoma at insertion sites, and checking peripheral pulses or limb perfusion where cannulas are placed.
63The complete ECMO system should be assessed at least daily for clots in the circuit, oxygenator, or pump head, for clicking or friction suggesting fibrin on the impeller, for pressure relationships across the oxygenator, and for cracks, loose connections, displacement, and battery backup.
64Clinical monitoring should include head-to-toe examination from drainage to reinfusion cannula each shift, looking for bleeding, infection, thrombosis, minimal color difference between inflow and outflow blood suggesting recirculation or absent sweep gas, chattering, membrane thrombosis or fibrin, pressure sores, and limb perfusion.
65Console monitoring includes ECMO mode, power connection, battery charge, flow, RPM, heat exchanger temperature, sweep gas flow and gas connection, membrane FiO2, and availability on the pump cart of a minimum of 6–8 tubing clamps, preferably guarded.
66Arterial blood gas measurement is recommended 1 hour after starting ECMO and then as clinically indicated.
67Plasma-free hemoglobin should be measured every day on ECMO, and trended daily as a tool to depict levels of hemolysis. A published threshold pair lists a plasma-free Hb of >0.1 g/dL as haemolysis and >1.0 g/L as severe haemolysis, but 0.1 g/dL equals 1.0 g/L (100 mg/dL), so the two printed thresholds are the same value and the pair is internally inconsistent. Plasma-free Hb >50 mg/dL is reported as the most common cutoff for clinically significant hemolysis and as the ELSO Registry definition. Severe haemolysis (plasma-free Hb >1.0 g/L, or with hyperkalaemia or red urine) suggests pump head thrombosis and must be treated with an urgent circuit change.
68Monitoring the distal limbs early with near-infrared spectroscopy, together with bedside Doppler and clinical examination, can pick up malperfusion at an early stage and permit timely intervention when needed. Lower-extremity perfusion should also be monitored continuously with NIRS, and NIRS probes may be placed on both legs before cannulation to assess perfusion and help decide whether to place a distal perfusion catheter (DPC). NIRS is a non-invasive modality that makes it possible to assess whether cerebral oxygen delivery is adequate, and trends and asymmetry between sides can indicate important changes in cerebral perfusion, enabling timely intervention when differential oxygenation is possible or when extremity malperfusion must be detected in peripheral ECLS. In pediatric respiratory ECLS patient management, near-infrared spectroscopy plays an important role in managing ECMO patients, especially with paired cerebral and somatic sensors, and the sensors promptly register reduced brain and kidney oxygenation or perfusion.