Cautions and pitfalls1ECLS is resource-intensive and has a high complication burden, so the decision to initiate it should weigh benefits against known risks and expected short- and long-term outcomes.
2ECLS should not be initiated when there is no hope of achieving a desired bridge outcome.
3For adult respiratory failure, the only absolute contraindication is an irreversible underlying condition with anticipated non-recovery in a patient who is not eligible for lung transplantation.
4Relative concerns include irreversible neurologic injury, moribund status, untreatable metastatic cancer, uncontrollable systemic bleeding especially in the CNS, contraindications to anticoagulation, prolonged mechanical ventilation with high airway pressures and FiO2, and advanced age.
5A separate ARDS table lists relative contraindications as invasive mechanical ventilation for >7-10 days, contraindication to anticoagulation, and severe coagulopathy, and absolute contraindications as moribund state with established multiple organ failure, prolonged cardiac arrest, severe anoxic brain injury, massive intracranial hemorrhage, severe chronic respiratory failure with no possibility of lung transplantation, and metastatic malignancy or hematological disease with poor short-term prognosis.
6There are no universally agreed absolute contraindications except end-stage respiratory failure when lung transplantation will not be considered.
7For neonatal respiratory ECLS, some historical absolute contraindications are now treated as relative contraindications and candidacy is fluid as technology and medical therapy advance.
8In CDH, relative contraindications listed include significant congenital anomalies, lethal chromosomal abnormalities such as Trisomy 18 or other lethal malformations, grade III/IV intracranial hemorrhage, weight <1.8 kg, gestational age <32-34 weeks, and prolonged mechanical ventilation with prolonged high pressure defined as MAP >15-18 x 7 days; ELSO Registry data report mortality of 71% with birth weight <2 kg and 56.3% with gestational age <34 weeks.
9Mortality is 71% for BW <2 kg and 56.3% for GA <34 weeks in registry data.
10A pediatric cardiac ECLS table lists poor primary disease prognosis, established multiorgan failure, severe pre-ECMO neurologic injury or intracranial hemorrhage, severe surgical or visceral bleeding, prematurity <34 w gestation, size <2 kg, and family preference as contraindications.
11Bleeding, thrombosis, and hemolysis are major complications in neonatal respiratory ECLS; in the BATE study, bleeding occurred in 60%, thrombosis in 44%, hemolysis in 40%, circuit thrombosis in 40% including entire circuit thrombosis in 28%, cannula-site bleeding in 26%, and surgical-site bleeding in 25%.
12Neurologic complications carry major prognostic importance: a systematic review of adult ECLS reported median acute neurologic complication risk 13%, including intracranial hemorrhage 5%, ischemic stroke 5%, and seizures 2%, with mortality 83% versus 42% in those without neurologic complications.
13In pediatric respiratory ECMO, intracranial hemorrhage occurred in 5%, associated with mortality of 79%; the ELSO registry figures give 6.4% in children, 7.6% in neonates, and 3.9% in adults.
14Baseline neurologic examination is recommended, and concerning findings include unstable hemodynamics, pupillary changes, seizures, new focal deficits, or abnormalities on neuromonitoring such as NIRS or EEG.
15In young children with an open fontanelle, cranial ultrasound can be used for surveillance; in one survey, 73% of pediatric centers used it routinely, with frequencies ranging from daily 25%, twice weekly 20%, three times weekly 17%, or only when clinically concerned 15%, and 74% would obtain brain CT if concern arose.
16The same survey reported NIRS as the most common neuromonitoring modality, used in 48% of adult centers and 80% of pediatric centers.
17Continuous surveillance of the ECLS system and circuit integrity and performance is mandatory, including visual inspection and performance parameters, with attention to oxygenator dysfunction or leak, circuit and vent-line thrombosis, and distal leg perfusion thrombosis.
18In adult post-cardiotomy ECLS, these lung-rest ventilator settings apply only in PCS patients receiving lung-rest ventilation to avoid high driving pressures when lung compliance is poor: plateau pressure <30 cm of water, PEEP 10–15 cm, pressure control 10 above PEEP, and respiratory rate 10.
19That same consensus also describes lung-protective targets of tidal volume <6 cc/kg, peak airway pressure <30 cm H2O, and avoidance of toxic FiO2 concentrations, with <40% cited as diminishing ARDS risk.
20Routine bronchoscopy is recommended to clear secretions and evaluate for infection, pulmonary hemorrhage, and atelectasis.
21Infections are frequent on ECLS; rates are 9-65% in post-cardiotomy V-A ECLS, with bloodstream infection 3-16%, lower respiratory tract infections 24.4 episodes/1000 days, and surgical site infection 0.6-14.7%.
22A low threshold for infection treatment is advised, and considering antifungal therapy is suggested in longer ECLS runs or in patients with chronic illness or immunodeficiencies.
23For thermal regulation devices, devices without patient blood temperature feedback can dangerously overheat or overcool patients, and strict cleaning and maintenance are emphasized because contaminated water recirculation devices were linked to Mycobacterium chimaera infections.
24Mode-specific complications differ: for VV ECMO they include bleeding including GI bleeding and intracranial hemorrhage, cannula malposition or dislodgement, oxygenator failure, ventilator-associated lung injury from high airway pressures, bloodstream infection from cannula sites, recirculation, ischemic stroke, and hemolysis.
25For VA ECMO they include limb ischemia due to distal reperfusion issues in femoral cannulation, aortic insufficiency/regurgitation due to retrograde flow, north-south syndrome, hemolysis, higher bleeding risk than VV ECMO, and air embolism.
26During VA ECMO, extracorporeal flow is ideally about 80% of total flow, corresponding in practice to a pulse pressure of about 10 mmHg.
27If the heart is completely nonfunctional on VA ECMO, left-sided filling pressure may rise to 20-25 mmHg and cause pulmonary edema and LV overdistension, and blood stasis in the chambers and pulmonary circulation can lead to clotting.
28Echocardiography is emphasized for assessing residual lesions, ventricular function, tamponade, inadequate decompression, spontaneous contrast, closed aortic valve, and cannula position, and more frequent studies may guide adjustments in flow, inotropes, and anticoagulation.
29Non-aggressive measures for LV distension are recommended first, including manipulating ECLS flow, vasodilation, and increased PEEP; if unresponsive, catheter-based LV unloading or septostomy is recommended.
30In the setting of bleeding, anticoagulation withdrawal may be considered with monitoring of oxygenator and pump circuit clot burden and function, cerebral emboli, and LV stasis and clot.